CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity
on circuits or wires that have voltage on them.
1. Set the function switch to the green
position.
2. Insert the black lead banana plug
into the negative COM jack.
Insert the red test lead banana plug
into the positive jack.
3. Press the SELECT button to
indicate on the display.
4. Touch the test probe tips to the
circuit or wire you wish to check.
5. If the resistance is less than
approximately 150, the audible
signal will sound. If the circuit is
open, the display will indicate “OL”.
DIODE TEST
1. Set the function switch to the
green position.
2. Insert the black test lead banana
plug into the negative COM jack
and the red test lead banana
plug into the positive diode jack.
3. Press the SELECT button to
indicate on the display.
4. Touch the test probes to the
diode under test. Forward
voltage will typically indicate
0.400 to 0.700V. Reverse
voltage will indicate “OL”..
Shorted devices will indicate
near 0V and an open device will
indicate “OL” in both polarities.