174-082010_Digital AVC Page 10 of 17
This signal is equal to an amplified tube millivolt signal. This signal will NOT be linearly proportional to
the indicated pressure. 1 volt (1.2 for DAVC4-1.2V) will correspond to a system pressure that is at least 1
order of magnitude less than the minimum detectable pressure. Increasing pressure will be indicated by a
decreasing voltage. The minimum detectable pressure is 0.2 millitorr for DAVC-5, 1 millitorr for DAVC-6
and 20 millitorr for DAVC-4.
The voltage signal can be mapped to a pressure value by using the following equation.
Where:
V = Voltage
P = pressure in Torr for DV4 & DV5 versions and millitorr for DV6
Parameters DV6 DV4
DAVC-4-1.2V
DV5
a -1623.22 -5.10184 -3.8115614 -0.25948
b -58.0442 -6.91233 -2.5905928 -42.23869
c -11732.2 -4.4943 -26.238798 -2.92598
d -130.397 -6.30995 -22.881611 -256.99510
e 13338.17 9.563177 24.483441 3.18016
2
2
1 dVbV
eVcVa
P
++
++
=