19
DL-00096 Rev. 7
Message
Batterylow
Fullycharged
Checkingpower
Nobattery
BattLifeNN%
Charging:NN%
Charging
RTXXXXX
BreathrateXX
Alarm cleared
Description
Estimatedbatterylifeless
than17minutes.Theunit
will also generate a high
priority alarm as described
above.
Displayedasthebattery
menu item when the
batteryis100%charged.
Displayedasthebattery
menu item when the
battery status has recently
changed and the unit
software is analyzing the
changeinstatus.
Displayedasthebattery
menu item when there are
no communications with
thebattery.
Displayedpercentageof
battery charge if at least
10%andthereisno
externalpowerconnected.
NN%displaysthecur-
rentbatterychargelevel.
Displayedwhenbattery
chargeisgreaterthan10%
butlessthan100%and
thereisexternalpower
connected.
Batterychargeislessthan
10%andthereisexternal
powerconnected.
Non-resettabletotalelapsed
timeindicator(hours).
Thepatient’saveragebreath
rate whenthedeviceis
deliveringthemaximum
amountofoxygenand
the bolusisreduced.Ifno
breaths are detected, the
most recent breath rate is
shown.
Apreviouslysetalarmhas
beenautomaticallycleared.
Action
Chargethebatterypack
by plugging in to power
supply.
Noactionisrequired.
Noactionisrequired.
Verifythatthebat-
tery pack is cor-
rectlyinstalled.Contact
CustomerServiceifthe
battery is fully inserted
and the message
continues to be displayed
longerthan30seconds.
Messageisdisplayed
when mode key is
pressed.
Messageisdisplayed
when mode key is
pressed.
Messageisdisplayed
when mode key is
pressed.
Noactionrequired.
Reduceactivitylevel.Be
sureairinlet/outlethas
notbeenblocked.
Note: The message will
automatically clear when
the device returns to normal
operation.
Noactionrequired.
Other Messages, continued
Note: The message will be
automatically cleared when
plugged in to power supply.
DL-00096 rev 7.indd 19 6/1/2009 4:48:34 PM