A SERVICE OF

logo

6 PRC30-en-EU_V1.3 7/14
STORING VALUES INTO MEMORY
There are 5 memory locations for each function. The default values stored in the locations are:
Memory
Location
Type
J, K, C, R, S, N
Type
T
Type
E
T/C mV
M1
0.0
o
C
(32.0
o
F)
0.0
o
C
(32.0
o
F)
0.0
o
C
(32.0
o
F)
0mV
M2
100.0
o
C
(212.0
o
F)
100.0
o
C
(212.0
o
F)
100.0
o
C
(212.0
o
F)
10.00mV
M3
500.0
o
C
(932.0
o
F)
200.0
o
C
(392.0
o
F)
200.0
o
C
(392.0
o
F)
25.00mV
M4
750.0
o
C
(1382.0
o
F)
300.0
o
C
(572.0
o
F),
500.0
o
C
(932.0
o
F)
40.00mV
M5
1000.0
o
C
(1832.0
o
F)
400.0
o
C
(752.0
o
F)
750.0
o
C
(1382.0
o
F)
50.00mV
Memory
Location
mA % mV V
M1
4.00mA 0.0% 0mV 0V
M2
8.00mA 25% 500mV 5V
M3
12.00mA 50% 1000mV 10V
M4
16.00mA 75% 1500mV 15V
M5
20.00mA 100% 2000mV 20V
To change the values in memory:
1. Select the SOURCE mode
2. Press the SET button to select the memory location to be changed.
3. Press the arrow buttons to adjust to the new value
4. Momentarily press the MEM button to store the value. The memory location icon will blink
while the value is being stored..