A SERVICE OF

logo

7 PRC30-en-EU_V1.3 7/14
Modes of Operation
Temperature
MEASURE (Input) Mode of Operation
1. Turn the meter ON.
2. “MEASURE” will appear in the display.
3. Press and Hold the MODE button for 1 second to select the temperature function.
4. Press and Hold the TYPE button to select the thermocouple type.
5. Momentarily press the UNIT button to select °F or °C.
6. Connect the thermocouple to the meter.
7. Read the measurement on the display.
Note: See the keypad description section to turn CJC on/off.
SOURCE (Output) Mode of Operation
In this mode, the unit can source the equivalent mV value for the temperature and the
thermocouple type selected. The values can be output either manually or stepped from memory
as described earlier.
1. Turn the meter ON
2. MEASURE” will appear in the display.
3. Press and Hold the MODE button to select the temperature function.
4. Press and Hold the TYPE button to select the thermocouple type.
5. Momentarily press the UNIT button to select °F or °C.
6. Momentarily press the “I/O” button to select SOURCE.
7. Connect the Calibration Cable from the meter to the device to be calibrated.
8. Use the and
buttons to set the desired output value in the lower display. The upper
display indicates the actual temperature or voltage value being sourced. If the upper display
does not match the set value, check the batteries or the connections to the device being
calibrated.