Description Command Example
Response
Format
Response
Example Range
Set probe type p[robe]=IT[S-90]/
IE[C751]/ C[VD]/
Y[SI400]/ T[HR]
pr ITS-90,
IEC751,
CVD, YSI400,
THERM
Read ITS-90 R(.01) or CVD R0 r[0] r0 r0: 9999.999 r0: 100.035
Set ITS-90 R(.01) or CVD R0 r[0]=n r0=100.106 1 to 900
Read ITS-90 coefficient a/b/c/d/a4/b4 a {a/b/c/d/a4/b4}:
-9.999999E-99
a: -2.8644101E-5
Set ITS-90 coefficient a/b/c/d/a4/b4=n a=-2.8644E-5 1.0 to 9.9e-18
Read CVD coefficient al[pha]/de[lta]/be[ta] al {al/de/be}: 9.99999999 al: 0.0038123
Set CVD coefficient al[pha]/de[lta]/be[ta]=n al=0.0038512 E+/-18
Read THERM thermistor
coefficient
b0/b1/b2/b3 b3 {b0/b1/b2/b3}:
9.999999e-99
b3: -9.944070e-12
Set THERM thermistor
coefficient
b0/b1/b2/b3=n b3=-9.944E-12 E+/-18
Read probe wires wi[res] wi wi: 9 wi: 4
Set probe wires wi[res]=2/3/4 wi=4 2,3,4
Read probe calibration date pc[al] pcal pc: yyyy-mm-dd pc: 1999-03-28
Set probe calibration date pc[al]=yyyy-mm-dd pc=1999-03-28 1999-01-01 to
2100-12-31
Read probe due date pd[ue] pdue pd: yyyy-mm-dd pd: 1999-03-28
Set probe due date pd[ue]=yyyy-mm-dd pd=1999-03-28 1999-01-01 to
2100-12-31
Read filter fi[lter] fi fi: 99.99999 fi: 0.5
Set filter fi[lter]=n fi=0 0.0 to 60.0
Read low range calibration *c1 *c1 c1: 99.999 c1: -0.0121
Set low range calibration *c1=n *c1=-0.012 -1.999 to 1.999
Read high range calibration *c2 *c2 c2: 9999.9 c2: -0.9
Set high range calibration *c2=n *c2=-0.9 -1980.0 to
1980.0
Read meter calibration date mc[al] mcal mc: yyyy-mm-dd mc: 1999-03-28
Set meter calibration date mc[al]=yyyy-mm-dd mc=1999-03-28 1999-01-01 to
2100-12-31
Read meter due date md[ue] mdue md: yyyy-mm-dd : 1999-03-28
Set meter due date md[ue]=yyyy-mm-dd md=1999-03-28 1999-01-01 to
2100-12-31
Legend: [] Optional Command data
{} Returns either information
1521 Handheld Thermometer Readout
User’s Guide
40