A SERVICE OF

logo

Safety Information: Read First
5
Warning
To avoid electrical shock or fire if a test tool
input is connected to more than 42 V peak (30
Vrms) or on circuits of more than 4800 VA:
Use only insulated voltage probes, test leads
and adapters supplied with the test tool, or
indicated by Fluke as suitable for the Fluke190
ScopeMeter series.
Before use, inspect voltage probes, test leads
and accessories for mechanical damage and
replace when damaged.
Remove all probes, test leads and accessories
that are not in use.
Always connect the battery charger first to the
ac outlet before connecting it to the test tool.
Do not connect the ground spring (figure 1,
item f) to voltages higher than 42 V peak (30
Vrms) from earth ground.
Do not apply voltages that differ more than 600
V from earth ground to any input when
measuring in a CAT III environment.
Do not apply voltages that differ more than
1000 V from earth ground to any input when
measuring in a CAT II environment.
Do not apply voltages that differ more than 600
V from each other to the isolated inputs when
measuring in a CAT III environment.
Do not apply voltages that differ more than
1000 V from each other to the isolated inputs
when measuring in a CAT II environment.
Do not apply input voltages above the rating of
the instrument. Use caution when using 1:1
test leads because the probe tip voltage will be
directly transmitted to the test tool.
Do not use exposed metal BNC or banana plug
connectors.
Do not insert metal objects into connectors.
Always use the test tool only in the manner
specified.
Voltage ratings that are mentioned in the warnings, are
given as limits for “working voltage”. They represent
V ac rms (50-60 Hz) for ac sinewave applications and as
V dc for dc applications.
Overvoltage Category III refers to distribution level and
fixed installation circuits inside a building.
Overvoltage Category II refers to local level, which is
applicable for appliances and portable equipment.