Instruction Manual Mi455 FREE & TOTAL SULPHUR DIOXIDE Minititrator for wine analysis
1212
1212
12
Open refill cap on the ORP electrode and
immerse the electrode approximately 2 cm
(0.8”) into the sample to be tested. Do not
touch stir bar with the tip of the electrode.
Ensure there is titrant in the dosing tip. If
not, hold the dosing tip over an empty
container and press PUMP for a few seconds
until the titrant is in the dosing tip.
Add the contents of one powder packet of
Mi555-006 Stabilizer into the beaker.
Press CAL button. The meter enters in pump
calibration mode.
Press START STOP in order to start the
system calibration.
At the end of the calibration, DONE appears
for a few seconds and the meter returns to
the measurement mode.