![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/0/ab/0ab6d7c6-af3c-4785-a11b-7ea2348026f7/0ab6d7c6-af3c-4785-a11b-7ea2348026f7-bg13.png)
ar345 Program Manual 1 9
____________________________________________________________________________________
.8 Starting Data Collection
Data collection can be started by pressing the "Go"or"Go/Erase first" button in th
ar345−Change Parameters window. It is a good idea to always let the plate clean fir
before actually starting the data collection.
In Multiple Data Sets mode, before actually sending the command to the scanner th
ar345−Run Parameters window will pop up. This window gives an overview ofthe current
programmed data sets. The buttons in the lower part of the window come with differen
layouts:
_______________________________________________________________________
ode Left Center Right
________________________________________________________________________
efore data "Go" "Change" "Close"
ollection really start reedit parameters Close window
uring data "Abort NOW" "Stop AFTER IMAGE" Close
ollection Aborts exposure Stops after current image Close window
________________________________________________________________________
To start data collection you will have to press the "Go" button in this window. During dat
collection the mar345−Run Parameters window monitors the progress of the data collection
The window can be opened by selecting the "Progress" option in the "Windows" submen
(see 3.2.1). Active data sets are highlighted. Start−ϕ and Image will be updated. The numbe
of oscillations is given by the multiplier in the ∆ϕfield. If the scanner is idle, this windo
cannot be obtained.
.9 Stopping Data Collection
Data collection can be stopped by pressing the "Stop" button in the main window or b
pressing the "Stop"or"Abort" button in the mar345−Run Parameters window. If th
scanner is scanning or erasing the imaging plate, this cycle will always be finished. Pressin
"Abort" in this moment will be the same as pressing "Stop". Otherwise, "Abort" closes th
X−ray beam shutter immediately and stops ϕ or distance movement.