The instrument serial port ground connection is referenced to the instrument's internal frame and safety
ground. Before connecting the communication cable to the instrument serial port, apply power to the
instrument and the host computer. Check the ground potential (voltage) between the serial port ground
pins on the computer and the instrument. The ground connection is pin 5 on DB-9 connectors and pin 7
on DB-25 connectors. Voltages in excess of 5VAC at 110V input power, or in excess of 10VAC at
220V input power may indicate a difference in ground wiring and can damage the instrument and/or the
computer. Check the wiring and take other steps as needed to reduce this difference before connecting
the communications cable. You may also use a data optoisolator.
Instrument Performance
Spectral Data
Range (nm) 400 - 700
Photometric Range 0 - 150%
Spectral Resolution 10 nm
Spectral Bandpass < 12 nm
Measurement Speed (at 23° C)
Warm-up time 1 second
Command-to-data output 1 second
Read-to-read 3 seconds
Warm-up Period 30 minutes prior to testing
Sample Storage Capacity 500 spectral samples
999 tristimulus samples
99 standards
99 setups
You may test the repeatability of your instrument as follows:
1. Turn the ColorFlex on and allow it to warm it up for 30 minutes. Meanwhile, clean the black glass
(45°/0° instruments) and white tile as described in Chapter 24 and allow the tiles to return to room
temperature.
2. Standardize the instrument.
3. Configure your color data display to show DE* using D65/10°.
4. Center the instrument’s white standard tile at the measurement port. Do not move or remove the tile
for the remainder of the test.
5. Read the white tile as a standard.
6. Configure the timed read feature to automatically make and average 20 more measurements with a
read interval of 7 seconds and then initiate reading of the tile as a sample.