A SERVICE OF

logo

12
<Setup procedure>
This section describes the procedure for installing the RANGE7/5 properly, and completing the preparations for measurement.
Make preparations according to the following procedures:
1) Installing the RANGE7/5.
2) Set the RANGE7/5 at a position relative to the proper target, and mount the lens suitable in question as the target of measure.
(
No
replacement lens for RANGE5.
)
3) Connect the AC adaptor to the RANGE7/5.
4) Connect the RANGE7/5 to a PC.
5) Turn ON the RANGE7/5 power switch to start warming up the RANGE7/5.
1) Installing the RANGE7/5
The RANGE7
/5
can be directly installed on a table by using the rubber soles at the bottom. Also, it can be installed with the tripod set or
measuring stand (under optional accessories). When the tripod set or measuring stand are used, the RANGE7
/5
orientation can be freely
changed (up/down and right/left). This feature is useful for measurement.
Memo
For installation procedure details, please refer to the description on “Installation”.
2)
Setting the RANGE7
/5
’s position relative to the measuring target, and exchanging the lens
(RANGE7 only)
Scan Position Adjustment
ThelocationandpostureoftheworkandRANGE7/5canbeadjustedsothatthetwoareappropriatelypositionedfor
scanning.Theseadjustmentsaremadebyactivatingthemonitoringfeatureandwatchingtheworkinthemonitorwindow.
Clickthe[Monitor]buttontochecktheworkonthemonitor
window.
Distancetoworkandworksize (Unit:mm)
Lens TELElens WIDElens
Distance 450 800 450 800
X×Y 79×99 141×176 150×188 267×334
Z 54 97 109 194
Memo
ForRANGE5,pleaserefertotheWIDElensdata.
Distancetoworkandworksize
in MultiFocus mode (Unit:mm)
Lens TELElens WIDElens
Distance 462 781 475 766
X×Y 81×102 138×172 159×199 256×320
Z 54 97 109 194
Memo
ForRANGE5,pleaserefertotheWIDElensdata.
Distancetowork
450mmto800mm
Receiver lens
RANGE7/5
Work
Y
X
Z
The origin is set at the sensor position which is
approximately 120 mm behind the center of the lens
surface. Therefore, the Z dimension has an offset of
around 120 mm applied.
RANGE7:TELE/WIDEreplaceable
RANGE5:Noreplacementlens
RANGE VIEWER monitor window