n Specications
MODELNAME KONICAMINOLTARANGE7
Measuring Method Triangulation by light sectioning method
Light-Source Semiconductorlaserλ=660nm
LaserClassication Class2(IEC60825-1edition2)
Sensor
Name CMOSPhotodiode
Pixel 1.31MegaPixel(1280x1024)
MeasuringDistance(mm) 450to800(DistancefromLight-ReceivingLens*1)
LightReceivingLens (Replaceable) TELE/WIDE
Measuring
Range
(mm)
MultiFocus
mode
Disabled
Lens TELE TELE WIDE WIDE
Distance 450 800 450 800
X-Ysize 79×99 141×176 150×188 267×334
Z 54 97 109 194
Meas.pitchofX-Y 0.08 0.14 0.16 0.28
MultiFocus
mode
Enabled
Lens TELE TELE WIDE WIDE
Distance 462 781 475 766
X-Ysize 81×102 138×172 159×199 256×320
Z 54 97 109 194
Meas.pitchofX-Y 0.08 0.13 0.08 0.25
Accuracy(Interglobulerdistance)*2 ±40μm
Precision(Z,σ)*3 4μm
AutoFocus available
AutoExposure available
ScanningTime about 2 sec. per scan
PreviewFunction available(about0.4sec.perscan)
EnvironmentBrightness 500lxorless
FileFormat*4 INPUT/OUTPUT:.rgv,.rvm,.rmk(Originalformat),OUTonly:.stl
OutputInterface USB2.0HighSpeed
PowerSupply
ACadapter
100-240Vac(50-60Hz),1.4A
ExternalDimensions(mm)
295(W)x190(H)x200(D)
* exclude the protruding portion of the lens and handle
Weight(kg) about6.7
OperatingCondition 10to40ºC(relativehumidity65%orless,withnocondensation)
StorageCondition
-10to50ºC(relativehumidity85%orless[at35ºC],withno
condensation)
Remarks(ControlledMethod) ControlledviadedicatedPCsoftware"RANGEVIEWER"
*1 Due to the difference of origins on the measurement reference surface and measurement data, the Z axis coordinate has an offset of about 120 mm for the measuring
distance.
*2 When the interglobuler distance for the ball bars (2 globes) specified in VDI/VDE2634-2 is measured under the following conditions defined by KONICA
MINOLTA OPTICS:
Our conditions: Temperature 20±1ºC / Using the TELE lens / Distance 450 mm / Warming up 20 mins. / Using the KONICA MINOLTA OPTICS processing
software / With Calibration / Object to be measured: Standard KONICA MINOLTA OPTICS (2 globes) instrument / Position of object to be
measured: KONICA MINOLTA OPTICS standard position (10 locations within the measurement space) / Does not cover the uncertainty over the
pricing of the standard instrument.
*3 Measuring conditions: Temperature 20±1ºC / Using the TELE lens / Distance 450 mm / Warming up 20 mins. / Using the KONICA MINOLTA OPTICS processing
software / Object to be measured: KONICA MINOLTA OPTICS reference plain chart / 1σ
*4 Using the KONICA MINOLTA OPTICS “RANGE VIEWER” processing software
• The RANGE7 includes the eT-Kernel/Compact from sSOL Co., Ltd.