Scanner Installation
Checking the Installation Site
2-3
Checking the Installation Site
Before you begin installing the OpScan iNSIGHT 70 scanner, verify the following
site conditions:
• Check the physical dimensions of the selected site to ensure adequate space.
– The OpScan iNSIGHT 70 scanner is 57.5 inches high, 48 inches wide, and
29 inches deep.
– If the scanner is near a wall, be sure to leave approximately 24 inches of
space between the scanner and the wall. This space is required so that an
operator can open the rear panel doors.
• Check that the site selected for the scanner meets all power and
environmental requirements.
– The temperature must be between 60
o
and 80
o
Fahrenheit (16 - 27 C).
– The humidity must be between 40% and 60% (non-condensing).
• Heat dissipation is 5300 - 5500 BTU/hr.
• Check that the circuit from the power source is a dedicated circuit for the
scanner.