Specifications
M3500B HeartStart XLT Defibrillator/Monitor 13-9
13
Environmental
Temperature: 0
o
to 50
o
C operating, -20
o
to 70
o
C storage.
Thermal paper may darken above 55
o
C.
Charging the battery at temperatures above 35
o
C may degrade battery
life.
Storing the battery for extended periods at temperatures above 40
o
C
will reduce battery capacity and degrade battery life.
Operating and storage specifications for electrodes may vary. Refer to
the manufacturer’s specifications for details.
Humidity:
Up to 95% Relative Humidity
Printer paper may jam if paper is wet.
Thermal printer may be damaged if wet paper is allowed to dry while
in contact with printer elements.
Altitude:
Operating: up to 15,000 ft.
Shipping: up to 20,000 ft.
Shock (drops onto concrete): Unit survives 39" (1 m) drops onto all surfaces
(faces, corners, and edges) enclosed in carrying case. Exposed handle surfaces
survive 24" drops.
Vibration: Mil Std 810E 514.4 Category 6 Helicopter, General Storage,
UH60.
Water Resistance: IPX4, splash proof per IEC 60529, with Data Card door
closed.
EMC: Meets EN 60601-1-2. This ISM device complies with Canadian
ICES-001.
Other Considerations: Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: Continuous.