Safety Considerations
M3500B HeartStart XLT Defibrillator/Monitor 13-21
13
WARNIN G Avoid contact between the patient and conductive fluids and/or metal objects, such
as the gurney. Contact with metal objects could cause unintentional current path-
ways.
WARNIN G Operating the HeartStart XLT or its accessories in conditions outside the environ-
mental specifications can result in device or accessory malfunction.
WARNIN G Avoid exposing the display to direct sunlight on a hot day. Overheating can occur and
cause the display to black out, making the HeartStart XLT temporarily unusable.
WARNIN G Medical electrical equipment which does not incorporate defibrillator protection
should be disconnected during defibrillation.
WARNIN G Electric shock hazards exist internally. Do not remove assembly screws. Refer ser-
vicing to qualified personnel.
CAUTION This device has not been evaluated for use with electrosurgery equipment.