Chapter 9 Troubleshooting
9-12 Applied Biosystems
9
Table 9-7 Calibration TroubleshootingMariner Only
Symptom Possible Cause Action
Mass Calibration
commands are dimmed
Chromatogram window is
selected
Select Spectrum window.
Apply Calibration
command is dimmed
when calibrating
MS Method data
For MS Method data,
calibration is valid for an
individual spectrum. You
cannot apply the
calibration from one
spectrum to the entire
data file.
No action. Normal
occurrence.
Error displayed when you
import a calibration
.CAL file corrupted Create new .CAL file. See
Exporting .BIC, .MSM,
and .CAL files on
page 1-36.
Importing a .CAL file
generated from a Voyager
data file
Import a Mariner .CAL.