After confirmation, the selected parameter
is displayed. The user can scroll through
the parameters by pressing CFM.
In order to directly set another param-
eter, press SETUP again and enter the
code or scroll to it by pressing CFM.
The following table lists the setup codes along with the descrip-
tion of the specific setup items, their valid values and whether
password is required to view that item (“PW” column):
Code Valid Values Default PW
00 Factory ID 0 to 9999 0000 no
01 Process ID 0 to 9999 0000 no
02 Control enable/disable 0: C.M. disabled 0 no
1: C.M. enabled
03 Range 1: 0.0-199.9 µS (or 100.0 ppm) 4 no
(depends on model) 2: 0-1999 µS (or 1000 ppm)
3: 0.00-19.99 mS (or 10.00 ppt)
4: 0.0-199.9 mS (or 100.0 ppt)
04 Reference Temperature 20°C or 25°C 25°C no
05 Temperature Coefficient 0.00 to 10.00 %/°C 2.00 no
06 Input Selection 0: conductivity probe 0 no
1: 4-20 mA input signal
07 Temperature ATC: Automatic ATC no
compensation User: Manual
08 TDS Factor (HI 710 only) 0.00 to 1.00 0.50 no
11 Relay 1 mode 0: disabled 0 no
(M1) 1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: PID, high setpoint
4: PID, low setpoint
12 Relay 1 setpoint (S1) 0.5 to 99.5% full scale 25% f.s. no
13 Relay 1 hysteresis (H1) 0 to 5% f.s. 1% f.s. no