36
LAST CALIBRATION DATA
The meter can display the following last calibration data:
• Date
• Time
• Cell constant
While displaying these data, the controller remains in control
mode. The data are related to the selected range only.
The procedure below indicates the flow. Displaying of the
items follows the above sequence.
• To begin the cycle press CAL DATA. The last calibration
date will appear on the primary LCD as DD.MM format,
while the secondary display will show the year.
If the meter has never been calibrated or an EEPROM reset
has occurred, no calibration data are shown when CAL
DATA is pressed. The “no CAL” message will blink for a few
seconds, then the meter skips back to normal mode.
• Press or to cycle through
the data forward or backwards
respectively.
Note In any moment, by pressing LCD or CAL DATA the meter will
return to the regular operating display.
• Press or to view the time of last
calibration. The secondary display will
show "HOU".
• Press or again to view the cell
constant at the time of last calibra-
tion. The secondary display will show
"CEL".
• Press or again to return to the first CAL DATA display
(date) at the time of last calibration.