A SERVICE OF

logo

9
Mode
Units
Setup Print
Enter
Mode
Units
Setup Print
Enter
3.3.1 Internal Calibration (InCAL
TM
)
On Explorer balances equipped with the InCal
TM
feature, calibration can be accomplished using the internal cali-
bration mass. When the balance requires calibration, a screen prompt of CAL NOW appears. This display can be
turned off as described in paragraph 3.3.2. Also, a software adjust feature is incorporated which permits the
internal calibration mass to be adjusted to +100 counts. The adjust feature is described in paragraph 3.3.3. Internal
calibration can be performed at any time providing the balance has warmed up to operating temperature.
Stable
g
Setup
Use keys to change, press Enter
Change
Use keys to change, press Enter
Change
Setup
Setup
Change
Setup
Change
Setup
Use keys to change, press Enter
Change
Procedure
Press the Setup button, CAL is displayed.
Press Enter button, CAL TYPE is displayed.
Press Enter button, CAL TYPE InCAL is displayed.
Press Enter button, INCAL is displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan). The
balance automatically resumes calibration.
The internal mass is positioned several times during
calibration and then removed, then after a few seconds,
CAL SET is displayed indicating a successful calibration.
The display then returns to WEIGH mode.
IMPORTANT !
DO NOT DISTURB THE BALANCE DURING
CALIBRATION. IF THE MESSAGE UNSTBLE IS DIS-
PLAYED, THE BALANCE WAS UNABLE TO ACQUIRE
STABLE DATA DURING INTERNAL CALIBRATION.
THE BALANCE WILL CONTINUE TO PERFORM
INTERNAL CALIBRATION UNTIL READINGS STABI-
LIZE. THE BALANCE WILL THEN COMPLETE THE
INTERNAL CALIBRATION FUNCTION.
TO EXIT INTERNAL CALIBRATION MODE BEFORE
COMPLETION, PRESS ENTER OR SETUP BUTTONS.
STABILITY CAN BE AFFECTED BY TEMPERATURE,
AIR CURRENTS, VIBRATION, ETC...