18
Mode
Units
Setup Print
Enter
Mode
Units
Setup Print
Enter
3.5 Percent Weighing
Percent Weighing is
enabled only
when Percent is turned ON in the Mode submenu under Setup. Percent
weighing permits you to place a reference load on the balance, then view other loads as a percentage of the
reference. The load you place on the pan as a reference may be displayed as any percentage you select from 5 %
to 100 % (in 1 % increments). One hundred percent does not necessarily have to represent the reference load.
Subsequent loads, displayed as a percentage of the reference are limited only by the capacity of the balance.
The default setting is Reference 100 %. Refer to paragraph 4.8 to enable percent weighing.
Stable
%
Mode
Use keys to change, press Enter
Change
Mode
Use keys to change, press Enter
Change
%
Procedure
• Press the Mode button.
• Press or button until PERCENT is dis-
played.
• Press Enter button, PUT>PAN 100 % is displayed.
If a container is used, the balance can be tared at
this point. The % display momentarily blanks while
the balance is taring out.
• Put the reference load on the pan.
• Press or button and select reference
weight percentage (Percent Range 5 to 100). Hold
button down for fast change.
• Press Enter button to save setting, WORKING is
displayed ... calculating reference weight.
Balance displays reference weight for five seconds in
selected measuring unit, then displays the percentage.
• Remove the reference weight from the pan and
replace it with another load. The second load is dis-
played as a percentage of the reference.
NOTE: The PERCENT display (number of digits) is a
function of the accuracy of the balance and the size of
the reference weight. The display examples were with a
300 g mass used with an 8.1 kilogram balance.
Stable
g