3.3.8 Calibration GLP Printout
If any option in the GLP Set Menu is turned On, GLP automatically prints data after calibration is completed.
- - - - - SPAN CAL - - - - - -
12/01/97 1:00:00 PM
Bal Id 1234
Cal: 8000.00 g
Old: 8000.00 g
Dif: 0.00 g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - INCAL - - - - - -
12/01/97 1:00:00 PM
Bal Id 1234
Cal: 8000.00 g
Old: 8000.00 g
Dif: 0.00 g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
InCAL
TM
Calibration Printout
When performing InCAL
TM
calibration with all GLP options
turned on, a printout is automatically made after the
calibration is completed.
- - - - - LIN CAL - - - - - -
12/01/97 1:00:00 PM
Bal Id 1234
Cal: 8000.00 g
Old: 7999.08 g
Dif: 0.02 g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - CAL TEST - - - - - -
12/01/97 1:00:00 PM
Bal Id 1234
Cal: 8000.00 g
Act: 8000.02 g
Dif: 0.02 g
Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
Span Calibration Printout
When performing Span calibration with all GLP options
turned on, a printout is automatically made after the
calibration is completed.
Linearity Calibration Printout
When performing a Linearity calibration with all GLP
options turned on, a printout is automatically made after
the calibration is completed.
Calibration Test Printout
When performing a Calibration Test with all GLP options
turned on, a printout is automatically made after the cali-
bration is completed.