21
Mode
Units
Setup Print
Enter
3.8 Weigh Below
The Explorer balance is equipped with a weigh below
hook at the bottom of the balance. To use this feature,
remove power from the balance and remove the pro-
tective cover underneath the balance. See illustration
for location. The balance can be supported using lab
jacks or any other convienent method. Make sure the
balance is level and secure. Apply power and operate
the balance. Attach items to be weighed to the hook
underneath the balance.
BOTTOM OF BALANCE
WEIGH BELOW
HOOK LOCATION
REMOVE
PROTECTIVE
COVER.
WEIGH BELOW
HOOK IS INSIDE OF
BALANCE.
3.9 Printing Data
Printing data to an external computer or printer requires that the communications parameters in the Setup menu,
Print options and communication parameters be set first. Refer to page 25 Print menu settings and page 27 for
RS232 communication settings.
Procedure
• Press the Print button. Printing to an external printer
or computer will occur each time the Print button is
pressed unless autoprint feature is turned on in which
case printing can occur in a continuous fashion, at
specified intervals or each time a stable reading is
achieved.
Sample printout is shown below with time turned on.
Setup
Use keys to change, press Enter
Change
SAMPLE PRINTOUT
For a review of of printing samples, refer to Section 4
Setting Up Your Balance. What is printed is controlled
by the GLP Set Menu and the selection of GLP Cont. or
GLP Tare in the Print Menu.
12/01/97 12:01:37 AM
429.5 g