A SERVICE OF

logo

12
Mode
Units
Setup Print
Enter
Mode
Units
Setup Print
Enter
3.3.4 Span Calibration
Span calibration utilizes two calibration points, one at zero and the other at full span or incremental values starting
at 25 % of full capacity. As an example, an 8.1 kg balance will accept either 2 kg, 4 kg, 6 kg or 8 kg for span
calibration. Values which are below or in between will not be accepted and the balance will display its maximum
capacity. When LFT is set on, only full span calibration value can be used. Sample display illustrates an 8.1 kg
balance
Stable
g
Setup
Use keys to change, press Enter
Change
Setup
Use keys to change, press Enter
Change
Use keys to change, press Enter
Change
Setup
Setup
Use keys to change, press Enter
Change
Setup
press Enter
Change
g
Setup
Change
Procedure
Press the Setup button, CAL is displayed.
Press Enter button, CAL TYPE is displayed.
Press Enter button, CAL TYPE InCAL is displayed.
Press button to select SPAN calibration, CAL
TYPE SPAN is displayed.
Press Enter button, WORKING is displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan). The
balance automatically resumes calibration.
Display changes to PUT WT 8000 g. The displayed
weight is the full capacity of the balance.
Place specified calibration mass on pan.
NOTE: For an 8.1 kg balance, either 2 kg, 4 kg, 6 kg or
8 kg can be used for span calibration. The PUT WEIGHT
message indicates the calibration mass that is on the
pan.
Press Enter button, WORKING is displayed. After a
few seconds CAL SET is displayed, the display then
returns to WEIGH mode.
• Span calibration is completed.
Remove calibration mass from the pan.