Adjust gate R1 so that the singlet bead population is located in gate R1 (Figure 3a).
Figure 3a
Adjust the FL3 PMT so that the median intensity of the top FL3 bead population
is around 5000 (Figure 3b). Adjust gate R3 as necessary so that the dim FL3 bead
population is located in gate R3 (Figure 3b). Do not adjust the R2 gate.
Bright Beads (R2) Median: 5139.70
FL1 (Median) Median: 2.53
Figure 3b
Adjust the FL1 PMT so that the median of FL1 is approximately 2.0 – 2.5
(Figure 3b).
Adjust the FL2 PMT value so that the median of FL2 is approximately 2.0 - 2.5
(Figure 3c).
FL2 (Median) Median: 2.21
Figure 3c
Run Cytometer Setup Beads tube B to adjust the compensation settings for
FL2 – %FL1.