18
Power Quality Analyzer Model 3945-B
3.6 AC Current Probe Model SR193
(3945-B accuracy included)
NOTE: When installing probes, face the arrow on the probe in the direction of
the load.
Nominal Range: 1000AAC for f ≤1kHz
Measurement Range: 3A to 1200AACmax(I>1000Anotcontinuously)
NOTE: Currents <0.5A will be displayed as zero with this probe.
Probe Output Signal: 1mVAC/AAC
Maximum Clamping Diameter:2"(52mm)
Safety: EN 61010-2-032, Pollution Degree 2, 300V Cat. IV, 600V Cat. III,
Reference Conditions:
Ambient temperature 73°F(23°C)±5°F(3°C)
Humidity 20 to 75% of RH
Frequency 48 to 65Hz
Distortion factor <1% no DC current
Magneticeldofexternalorigin <40A/m(earth’smagneticeld)
Accuracy*
Primarycurrent(AAC) 3 to 10A 10 to 100A 100 to 1200A
Accuracy(%oftheoutputsignal)
≤0.8% ± 1ct ≤0.3% ± 1ct ≤0.2% ± 1ct
Phaseshift(°)
≤1° ≤0.5° ≤0.3°
* Logarithmic interpolation between each specied value
Factors affecting accuracy (% of the output signal)
Conditions Range Error
Temperature 14°to122°F(-10°to50°C)
≤200 ppm/°C or 0.2%
per 10°C
Relative humidity 10 to 90%
<0.1%
Frequency
30 to 48Hz
65 to 1000Hz
1 to 5kHz
<0.5%
<1%
<2%
Position of the cable in the jaws
– <0.1% @ ≤ 400Hz
Adjacent conductor carrying a
60Hz AC current
Conductor in contact
with the sensor
≤0.5mA/A
Crest factor distortion
≤6 and current ≤3000A peak <1%
DC current distortion
≤15ADC on the
nominal AC current
<1%
Overload: Frequencyderatingbeyond1kHz:
I max ≤ 1000A x
f (in kHz)
1