Power Quality Analyzer Model 3945-B
3.8 AC Current Probe MN193 Probe
(3945-B accuracy included)
NOTE: When installing probes, face the arrow on the probe in the direction of
the load.
The 5A range of the MN193 is designed to work with secondary current transform-
ers.Bestaccuracyisavailablewhenenteringthetransformerratio(e.g.1000/5A).
Whenusedtomeasure5Adirect,theresolutionwillbelimitedto0.1Amax.
Nominal Range: 5A and 100AAC
Measurement Range:
5A:0.005Ato6AACmax(1Ato1200Awithratio1000/5selected)
100A:0.1Ato120AACmax
NOTE: Currents < (Primary x 5) ÷ (Secondary x 1000) on the 5A range and <0.2A
on the 100A range will be displayed as zero with this probe
Probe Output Signal: 5A:200mV/AAC; 100A:10mV/AAC
Maximum Clamping Diameter:0.8"(20mm)
Safety: NF EN 61010-2-032, Pollution Degree 2, 300V Cat. IV, 600V Cat. III,
Reference Conditions:
Ambient temperature 73°F(23°C)±5°F(3°C)
Humidity 20 to 75% of RH
Frequency 48 to 65Hz
Distortion factor <1% without superimposed DC current
Magneticeldofexternalorigin <40A/m(earth’smagneticeld)
Accuracy Specications
Range: 10mV/A(1V@100Arms)
Primarycurrent(inAAC) 0.1A to 1A 1A to 120A
Accuracy(%oftheoutputsignal)
1% ± 2ct 1% ± 1ct
Phaseshift(°)
1.5°
Overload: 120A continuous
Range: 200mV/A(1V@5Arms)withratio1000/5selected
Primarycurrent(inAAC)
Probe output
1 to 10A
5 to 50mA
10 to 100A
0.05 to 0.5A
100 to 1200A
0.5 to 6A
Accuracy(%oftheoutputsignal)
1.5% ± 1ct 1.5% ± 1ct 1% ± 1ct
Phaseshift(°)
1.7°
Overload: 12A continuous