Microplate Luminometer Maintenance
104
8.11 Replacing Injector Tips
The injector tips are located in the injector head behind the
service door of the MONOLIGHT 3096.
Measurement head Injector head
Turn MONOLIGHT 3096 off and disconnect it from wall socket.
Open instrument door of MONOLIGHT 3096, open screws of
service door and take door off.
Open the 2 fixing screws of the injector head and take off the
injector head directly behind the measurement head.
Injector head Injector tubing Injector head (front view)
Injector tip
Unscrew the respective tubing fitting screw (made of metal)
with tubing and pull it out.
Push out injector tips towards the top.
Insert new injector tip and screw fitting screw into the injector
head (see also the previous section 8.10).
Figure 8-12:
Position of injector tip
in injector head
Figure 8-11:
MONOLIGHT 3096
open