Microplate Luminometer Maintenance
105
Insert injector head (the sloping part is facing you). Fit the in-
jector tips into the boreholes in the MONOLIGHT 3096 and the
adjusting bolts in the recesses in the injector head (Figure 8-
13).
Insert fixing screws and tighten them.
Close service door properly (light-tight) (see also page 20).
Boreholes for fixing screws Measurement head
Injector head (front view) Boreholes for injector tips Adjusting bolts
Figure 8-13:
Installing the
injector head