Microplate Luminometer Slow Kinetics
68
[Volume]: Enter the injection volumes in µl (via keyboard or by
clicking on the arrow buttons).
[Injection 1 to injection 2 delay]: Delay time between the injec-
tions of injector 1 and 2. The hardware-dependent minimum
value of 2.05 seconds is defaulted and can be increased (t
P-I
).
[Injection 2 to measurement delay]: Enter the delay time be-
tween the injection of injector 2 and the start of the measure-
ment in seconds. 0 second is defaulted and can be increased
(t
M-I
).
Example: Programming a measurement with 2 injections:
P = pre-injection
I= injection
M= measurement
t
P-I
= delay between pre-injection and 2
nd
injection
t
M-I
= delay between 2
nd
injection and start of measurement
T
M
= duration of measurement
Dual Measurement
If this item is checked, each selected sample is measured twice
and both measured values are displayed in the respective field.
This function can be used with or without injectors. Parameters
must be entered accordingly:
No injector: Delay time: 1
st
measurement -> 2
nd
measurement
Injector 1: Delay times: Inj. 1 -> 1
st
measurement -> 2
nd
measurement
Injector 2: Delay times: 1
st
measurement -> Inj. 2 -> 2
nd
measurement
Injector 1&2: Delay times: Inj. 1 -> 1
st
measurement -> Inj. 2 -> 2
nd
measurement
Injector 1 Injector 2
Microplate