Microplate Luminometer Maintenance
91
8.2 Fuse Replacement (MONOLIGHT 3096)
The fuses on the instrument rear panel must be replaced when
they are faulty or when changing the operating voltage (see 8.3).
The fuses are in a black fuse holder next to the power cord con-
nector (see Figure 8-1).
Power cord connector
0
100V
120V
230V
240V
I
Power switch Recess for screwdriver Fuse holder
Proceed as follows
You need:
1 small flat head screwdriver
Spare fuses:
For operating voltage 230 V: Fuse: 160 mA slow-blow
For operating voltage 115/120 V: Fuse: 315 mA slow-blow
Use only UL-approved fuses!
Turn power switch to off ("0").
Unplug the power cord from the wall outlet.
Insert a screwdriver or a similar tool into the recess on the
right side of the fuse holder, push slightly and take the fuse
holder out. You will see two fuses.
Take the faulty fuse(s) out. Note: You need both fuses for op-
eration. Faulty fuses can typically be detected by a broken
fuse wire in the glass envelope.
Figure 8-1:
Power supply unit
on the instrument
rear panel